• Title of article

    Laser ablation and ionization time-of-flight mass spectrometer with orthogonal sample introduction and axial field rf-only quadrupole cooling

  • Author/Authors

    He، نويسنده , , Jian and Zhong، نويسنده , , Wenwan and Mahan، نويسنده , , Cynthia and Hang، نويسنده , , Wei، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    220
  • To page
    224
  • Abstract
    We describe a laser ablation and ionization time-of-flight mass spectrometer with orthogonal sample introduction and axial field rf-only quadrupole cooling device. Experimental results have shown the fairly uniform sensitivities for different metal elements, and a satisfactory resolving power. Little interference is shown in the spectrum. The detection limits are about 10− 7–10− 8 g/g for most of the metal elements.
  • Keywords
    elemental analysis , mass spectrometry , Laser ionization , Direct solids analysis
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2006
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number