Title of article :
Limits of detection and decision. Part 3
Author/Authors :
Daniel and Voigtman، نويسنده , , E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
12
From page :
142
To page :
153
Abstract :
It has been shown that the MARLAP (Multi-Agency Radiological Laboratory Analytical Protocols) for estimating the Currie detection limit, which is based on ‘critical values of the non-centrality parameter of the non-central t distribution’, is intrinsically biased, even if no calibration curve or regression is used. This completed the refutation of the method, begun in Part 2. With the field cleared of obstructions, the true theory underlying Currieʹs limits of decision, detection and quantification, as they apply in a simple linear chemical measurement system (CMS) having heteroscedastic, Gaussian measurement noise and using weighted least squares (WLS) processing, was then derived. Extensive Monte Carlo simulations were performed, on 900 million independent calibration curves, for linear, “hockey stick” and quadratic noise precision models (NPMs). With errorless NPM parameters, all the simulation results were found to be in excellent agreement with the derived theoretical expressions. Even with as much as 30% noise on all of the relevant NPM parameters, the worst absolute errors in rates of false positives and false negatives, was only 0.3%.
Keywords :
bias , Limit of detection , detection limit , heteroscedastic
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2008
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1687201
Link To Document :
بازگشت