Title of article :
Femtosecond laser ablation: Experimental study of the repetition rate influence on inductively coupled plasma mass spectrometry performance
Author/Authors :
Gonzalez، نويسنده , , Jhanis J. and Fernandez، نويسنده , , Alberto and Oropeza، نويسنده , , Dayana and Mao، نويسنده , , Xianglei and Russo، نويسنده , , Richard E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
10
From page :
277
To page :
286
Abstract :
This paper demonstrates the feasibility of performing bulk chemical analysis based on laser ablation for good lateral resolution with only nominal mass ablated per pulse. The influence of repetition rate (1–1000 Hz) and scan speed (1–200 µm/s) using a low energy (30 µJ) and a small spot size (~ 10 µm) UV-femtosecond laser beam was evaluated for chemical analysis of silica glass samples, based on laser ablation sampling and inductively coupled plasma mass spectrometry (ICP-MS). Accuracy to approximately 14% and precision of 6% relative standard deviation (RSD) were measured.
Keywords :
Repetition rate effects , Femtosecond LA-ICP-MS , LA-ICP-MS
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2008
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1687222
Link To Document :
بازگشت