Title of article
Determination of Te in soldering tin using continuous flowing electrochemical hydride generation atomic fluorescence spectrometry
Author/Authors
Jiang، نويسنده , , Xianjuan and Gan، نويسنده , , Wuer and Han، نويسنده , , Suping and He، نويسنده , , Youzhao He، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
4
From page
710
To page
713
Abstract
An electrochemical hydride generation system was developed for the detection of Te by coupling an electrochemical hydride generator with atomic fluorescence spectrometry. Since TeH2 is unstable and easily decomposes in solution, a reticular W filament cathode was used in the present system. The TeH2 generated on the cathode surface was effectively driven out by sweeping gas from the cathode chamber. In addition, a low temperature electrochemical cell (10 °C) was applied to reduce the decomposition of TeH2 in solution. The limit of detection (LOD) was 2.2 ng ml− 1 and the relative standard deviation (RSD) was 3.9% for nine consecutive measurements of standard solution. This method was successfully employed for determination of Te in soldering tin material.
Keywords
Atomic fluorescence spectroscopy , TE , Electrochemical hydride generation
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2008
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1687312
Link To Document