• Title of article

    Polydisperse NiTi nanoparticles investigated by X-ray standing waves and electron microscopy—A comparative study

  • Author/Authors

    Brücher، نويسنده , , Martin and Chakif، نويسنده , , MʹBarek and Gurevich، نويسنده , , Evgeny L. and Hergenrِder، نويسنده , , Roland، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    5
  • From page
    60
  • To page
    64
  • Abstract
    A polydisperse mixture of nickel–titanium nanoparticles generated by femtosecond-laser ablation was investigated by the application of different analytical methods in order to characterize the distribution of particle sizes. Images obtained with scanning and transmission electron microscopy and intensity curves of fluorescence excited by X-ray standing waves (XSW) were evaluated and the resulting distributions were plotted in several histograms. Based on the differences found in the results, the possibilities, limitations and appropriate criteria of application of the respective technique are discussed. The principles of the XSW technique and the evaluation procedure are explained in more detail. The respective analytical methods were compared in terms of spatial resolution, information content, the risk of artifacts, the statistics of the evaluation and the availability of experimental facilities.
  • Keywords
    Polydisperse nanoparticles , Size distribution , X-ray standing waves , Fluorescence scan , Electron microscopy
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2014
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number

    1689176