Detecting electron beam energy shifts with a commercially available energy monitor
Author/Authors
Evans، Michael D. C. نويسنده , , Moftah، Belal A. نويسنده , , Olivares، Marina نويسنده , , Podgorsak، Ervin B. نويسنده ,
Issue Information
دوماهنامه با شماره پیاپی سال 2016
Pages
-1660
From page
1661
To page
0
Abstract
Routine electron beam quality assurance requires an accurate, yet practical, method of energy characterization. Subtle shifts in beam energy may be produced by the linac bending magnet assembly, and the sensitivity of a commercially available electron beam energy-monitoring device for monitoring these small energy drifts has been evaluated. The device shows an 11% change in signal for a 2 mm change in the I(50) energy parameter for low energy electron beams (in the vicinity of 6 MeV) and a 2.5% change in signal for a 2 mm change in the I(50) energy parameter for high energy electron beams (in the vicinity of 22 MeV). Thus the device is capable of detecting small energy shifts resulting from bending magnet drift for all clinically relevant electron beams.