Title of article
An investigation of the feasibility of gadolinium for neutron capture synovectomy
Author/Authors
Gierga، D. P. نويسنده , , Yanch، J. C. نويسنده , , Shefer، R. E. نويسنده ,
Issue Information
فصلنامه با شماره پیاپی سال 2017
Pages
-1684
From page
1685
To page
0
Abstract
Two patients developed skin ulcers secondary to high doses of diagnostic-energy x rays received during localization procedures as part of brachytherapy treatments. Both were morbidly obese and diabetic. The obesity led to the delivery of estimated skin doses of 83 Gy in one case and 29 Gy in the other in attempts to produce readable images on localization radiographs. This report discusses the factors leading to the injuries, the progression of the injuries over time. and the variables involved in the localization procedures with the aim of preventing future mishaps. The greatest contribution to the large skin dose was the need, with the equipment available, to use multiple exposures to produce a single film, because of the effect of the resultant reciprocity failure.
Keywords
gadolinium neutron capture synovectomy (GNCS), boron neutron capture therapy (BNCT) , neutron beam , accelerator , boron neutron capture synovectomy (BNCS) , rheumatoid arthritis