Title of article :
Determination of finasteride in human plasma by liquid–liquid extraction and high-performance liquid chromatography
Author/Authors :
Pt??ek، نويسنده , , P and Macek، نويسنده , , J and Kl??ma، نويسنده , , J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
6
From page :
305
To page :
310
Abstract :
A high-performance liquid chromatographic method for the quantitation of finasteride in human plasma is presented. The method is based on liquid–liquid extraction with hexane–isoamylalcohol (98:2, v/v) and reversed-phase chromatography with spectrophotometric detection at 210 nm. The mobile phase consists of acetonitrile–15 mM potassium dihydrogenphosphate (40:60, v/v). Clobazam is used as the internal standard. The limit of quantitation is 4 ng/ml and the calibration curve is linear up to 300 ng/ml. Within-day and between-day precision expressed by relative standard deviation is less than 5% and inaccuracy does not exceed 8%. The assay was used for pharmacokinetic studies.
Keywords :
Finasteride
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2000
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1702917