Title of article :
Rapid and sensitive high-performance liquid chromatographic method for busulfan assay in plasma
Author/Authors :
Nathalie Bleyzac، نويسنده , , Nathalie and Barou، نويسنده , , Philippe and Aulagner، نويسنده , , Gilles، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
6
From page :
427
To page :
432
Abstract :
A reversed-phase liquid chromatographic method with ultraviolet detection has been developed to determine busulfan concentrations in plasma of children undergoing bone marrow transplantation. Plasma samples (200 μl) containing busulfan and 1,6-bis(methanesulfonyloxy)hexane as an internal standard were prepared by a simple derivatization method with diethyldithiocarbamate followed by extraction with ethyl acetate and solid-phase purification on C8 columns conditioned with methanol and water and eluted with acetonitrile (recovery 99%). Chromatography was accomplished using a Hypersil octadecylsilyl column (10 cm×4.6 mm I.D.) and a mobile phase of acetonitrile, tetrahydrofuran and distilled water (65:5:30, v/v). The limit of detection was 25 ng/ml (signal-to-noise ratio of 5). Calibration curves were linear up to 25 000 ng/ml. Intra-day and inter-day coefficients of variation of the assay were ≤5%. This method was used to analyse busulfan plasma concentrations after oral administration within the framework of therapeutic drug monitoring and pharmacokinetic studies in children.
Keywords :
Busulfan
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2000
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1703453
Link To Document :
بازگشت