Title of article :
Automated solid-phase extraction method for the determination of atovaquone in plasma and whole blood by rapid high-performance liquid chromatography
Author/Authors :
Lindegهrdh، نويسنده , , N and Bergqvist، نويسنده , , Y، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
9
From page :
9
To page :
17
Abstract :
A bioanalytical method for the determination of atovaquone in plasma and whole blood by solid-phase extraction and high-performance liquid chromatography has been developed and validated. A structurally similar internal standard was added and protein was precipitated from plasma and whole blood with acetonitrile before being loaded on to a C8 solid-phase extraction column. Atovaquone and internal standard were analysed by high-performance liquid chromatography on a C18 J’Sphere ODS-M80 (150×4.0 mm) column with mobile phase acetonitrile–phosphate buffer, 0.01 M, pH 7.0 (65:35, v/v) and UV detection at 277 nm. The intra-assay precisions for plasma and whole blood were 2.2% and 1.9% respectively at 12 μM and 6.0% and 5.6% respectively at 0.75 μM. The inter-assay precisions for plasma and whole blood were 1.4% and 2.1% respectively at 12 μM and 4.9% and 3.4% respectively at 0.75 μM. The lower limit of quantification in plasma and whole blood were 150 nM. The limit of detection in plasma and whole blood were 30 nM.
Keywords :
Atovaquone
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2000
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1703605
Link To Document :
بازگشت