Title of article :
Trace determination of urinary 3-hydroxybenzo[a]pyrene by automated column-switching high-performance liquid chromatography
Author/Authors :
Simon، نويسنده , , P and Lafontaine، نويسنده , , M and Delsaut، نويسنده , , P and Morele، نويسنده , , Y and Nicot، نويسنده , , T، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
3-Hydroxybenzo[a]pyrene (3-OHB[a]P), one of the metabolites of benzo[a]pyrene (B[a]P), has been determined in human urine using an automated column-switching procedure. The hydrolysed biological sample is centrifuged just prior to being injected into a reusable precolumn loop, which is packed with a preparative phase and coupled on-line to a liquid chromatographic (LC) system. A rapid pre-treatment of the hydrolysed sample, consisting of a concentration and a crude clean-up, is performed on the precolumn. The analytes are then non-selectively desorbed with the LC eluent and the sample is cleaned again in three successive purification columns using the direct transfer or “heart-cut” technique. The pre-treatment does not exceed 3 min. and the entire analytical purification and separation procedure takes less than 30 min. Average 3-OHB[a]P recovery reaches 95% in the 1–50 ng/l range of urine, and the detection limit is 0.1 ng/l urine for a 3 ml injection of hydrolysed urine. The developed method was compared with a more time-consuming off-line method to analyse urines of B[a]P gavaged rats; the statistical treatment indicates that both methods are in agreement. The method was applied to purify and concentrate the urine samples of workers exposed and apparently unexposed to polycyclic aromatic hydrocarbons (PAHs).
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications