Abstract :
A system has been developed for the determination of quinolinate phosphoribosyltransferase (QPRT) activity in liver and kidney homogenates using HPLC. A product, nicotinic acid mononucleotide (NaMN), is separated by reversed-phase chromatography (a Tosoh ODS 80TS was used as an analytical column) using a mixture of 10 mM KH2PO4–K2HPO4 buffer (pH 7.0) containing 1.48 g/l tetra-n-butylammonium bromide–acetonitrile (9:1, v/v) as a mobile phase. The flow-rate was 1.0 ml/min, the detection wavelength was 265 nm. The column temperature was maintained at 40°C. Under these conditions, NaMN was eluted at about 8.1 min. Sample preparation was very straightforward. The reaction mixture of QPRT assay was stopped by immersing the tube into a boiling water bath, the resulting supernatant was filtered, and the filtrate was directly injected into a HPLC system. The total HPLC analysis time was approximately 20 min.
Keywords :