Title of article
Determination of scopolamine in human serum and microdialysis samples by liquid chromatography–tandem mass spectrometry
Author/Authors
Oertel، نويسنده , , R and Richter، نويسنده , , K and Ebert، نويسنده , , U and Kirch، نويسنده , , W، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
8
From page
121
To page
128
Abstract
A liquid chromatographic-tandem mass spectrometric (LC–MS–MS) method with a rapid and simple sample preparation was developed for the determination of scopolamine in biological fluids. Scopolamine and the internal standard atropine in serum samples were extracted and cleaned up by using an automated solid phase extraction method. Microdialysis samples were directly injected into the LC–MS system. The mass spectrometer was operated in the multi reaction monitoring mode. A good linear response over the range of 20 pg/ml to 5 ng/ml was demonstrated. The accuracy for added scopolamine ranged from 95.0 to 104.0%. The lower limit of quantification was 20 pg/ml. This method is suitable for pharmacokinetic studies.
Keywords
Scopolamine
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year
2001
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Record number
1704291