Title of article :
High-performance liquid chromatographic–tandem mass spectrometric determination of 3-hydroxypropylmercapturic acid in human urine
Author/Authors :
Mascher، نويسنده , , Daniel G and Mascher، نويسنده , , Hermann J and Scherer، نويسنده , , Gerhard and Schmid، نويسنده , , E.R، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
7
From page :
163
To page :
169
Abstract :
A sensitive and specific high-performance liquid chromatographic–tandem mass spectrometric (HPLC–MS–MS) method was developed for the determination of 3-hydroxypropylmercapturic acid (3-HPMA) in human urine. Samples were extracted using ENV+ cartridges and then injected onto a C8 Superspher Select B column with acetonitrile and formic acid as eluent (5:95, v/v). N-Acetylcysteine was used as internal standard for HPLC–MS–MS. Linearity was given in the tested range of 50–5000 ng/ml urine. The limit of quantification was 50 ng/ml. Precision, as C.V., in the tested range of 50–5000 ng/ml was 1.47–6.04%. Accuracy ranged from 87 to 114%. 3-HPMA was stable in human urine at 37°C for 24 h. The method was able to quantify 3-HPMA in urine of non-smokers and smokers.
Keywords :
3-Hydroxypropylmercapturic acid
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2001
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1704305
Link To Document :
بازگشت