Title of article :
A high-performance liquid chromatography–tandem mass spectrometric method for the determination of pharmacokinetics of ganaxolone in rat, monkey, dog and human plasma
Author/Authors :
Ramu، نويسنده , , Kumar and Lam، نويسنده , , Gilbert N and Chien، نويسنده , , Ben، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
11
From page :
49
To page :
59
Abstract :
A method for determining concentration levels of ganaxolone in rat, monkey, dog and human plasma was validated in the range of 5–1500 ng/ml using a 200-μl plasma sample volume. This validation report describes the linearity, specificity, sensitivity, reproducibility, accuracy, recovery and stability of the analytical method. The inter-day C.V. ranged from 0.5 to 9.2%, intra-day C.V. from 0.7 to 8.8% and intra-day accuracy (mean absolute percentage difference) ranged from 0.0 to 14.0% for rat, monkey, dog and human plasma. The method was used for the routine analysis of ganaxolone in rat, monkey, dog and human plasma and summary of the pharmacokinetic data are presented.
Keywords :
Ganaxolone
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2001
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1704327