Title of article :
Rapid determination of ranitidine in human plasma by high-performance liquid chromatography without solvent extraction
Author/Authors :
Ahmadiani، نويسنده , , Abolhassan and Amini، نويسنده , , Hossein، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
6
From page :
291
To page :
296
Abstract :
A simple high-performance liquid chromatographic procedure was developed for the determination of ranitidine in human plasma. The method entailed direct injection of the plasma samples after deproteination using perchloric acid. The chromatographic separation was accomplished with an isocratic elution using mobile phase consisting of 21 mM disodium hydrogen phosphate–triethylamine-acetonitrile (1000:60:150, v/v), pH 3.5. Analyses were run at a flow-rate of 1.3 ml/min using a μbondapak C18 column and ultraviolet detection at a wavelength of 320 nm. The method was specific and sensitive, with a quantification limit of approximately 20 ng/ml and a detection limit of 5 ng/ml at a signal-to-noise ratio of 3:1. The mean absolute recovery was about 96%, while the within- and between-day coefficient of variation and percent error values of the assay method were all less than 8%. The linearity was assessed in the range of 20–1000 ng/ml plasma, with a correlation coefficient of greater than 0.999. This method has been used to analyze several hundred human plasma samples for bioavailability studies.
Keywords :
Ranitidine
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2001
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1704420
Link To Document :
بازگشت