• Title of article

    Determination of moxifloxacin in growth media by high-performance liquid chromatography

  • Author/Authors

    Ba، نويسنده , , Boubakar B. and Etienne، نويسنده , , Renaud and Ducint، نويسنده , , Dominique and Quentin، نويسنده , , Claudine and Saux، نويسنده , , Marie-Claude، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    107
  • To page
    112
  • Abstract
    A direct injection high-performance liquid chromatographic method with column switching has been developed to determine moxifloxacin in Mueller–Hinton broth. A LiChrocart 4-4 pre-column filled with a LiChrospher 100 RP 18, 5 μm and a 150×4.6 mm I.D. column packed with a Supelcozil ABZ+ Plus were used and led to a retention time of 5.70 min. Fluorescence detection allowed one to reach a quantification limit of 0.05 μg/ml with a 100-μl sample size. The standard curves were linear from 0.05 to 3.2 μg/ml. Intra- and inter-day imprecisions within the linearity range were ≤4.76 and ≤5.75%, respectively. The mean relative errors for the same day and the day-to-day inaccuracies ranged from −2.93 to +4.50% and from −1.10 to +6.00%, respectively. The method was demonstrated to be useful for pharmacokinetic–pharmacodynamic studies of moxifloxacin in an in vitro model.
  • Keywords
    Moxifloxacin
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number