Title of article :
Improved method for the measurement of large neutral amino acids in biological matrices
Author/Authors :
Bongiovanni، نويسنده , , Rodolfo and Yamamoto، نويسنده , , Bryan K. and Jaskiw، نويسنده , , George E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
A high-performance liquid chromatographic method for measuring neutral amino acids in rat sera, brain tissues, and perfusates was developed by using o-phthalaldehyde sulfite as a pre-column derivatization reagent. With the present method, it was possible to separate the neutral amino acids within a single run in 25 min, while the acidic amino acids were eluted near or at the solvent front. The recovery was above 88.8% with a relative standard deviation (RSD) below 4.2%. The within- and between-day assay reproducibility for the determination of rat serum amino acids showed RSDs below 1.35 and 7.61%, respectively. In the present study, the neutral amino acids were assayed with high sensitivity, accuracy and good reproducibility in a relatively short time and on a small sample size.
Keywords :
amino acids , Neutral
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications