Title of article :
Automated sample preparation and gas chromatographic–mass spectrometric analysis of urinary androgenic anabolic steroids
Author/Authors :
Haber، نويسنده , , E and Muٌoz-Guerra، نويسنده , , J.A. and Soriano، نويسنده , , C and Carreras، نويسنده , , D and Rodriguez، نويسنده , , C and Rodriguez، نويسنده , , F.A، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
This paper presents an automated method for extracting anabolic agents from urine samples for their GC–MS analysis by selected-ion monitoring. The sample preparation was carried out in a Hewlett-Packard 7686 SPE PrepStation system. Each 0.6-ml aliquot was hydrolyzed, extracted, dried and trimethylsilyl (TMS) derivatized in a 2-ml vial without any hands-on labor. When sample preparation was finished 2 μl of the extract was injected into the gas chromatograph by split (1:10) mode. Due to the small amount of free space in the 2-ml vials for handling the sample, parameters like time of hydrolysis, type of shaking, number of extractions and some TMS derivatization parameters had to be adjusted to achieve the best recovery for all of the compounds in the screening. Manual and automated sample preparation schemes were compared in terms of linearity, precision, accuracy, limit of detection and recovery data. When large concentrations were analyzed using the automated method no carry-over effect was observed.
Keywords :
Anabolic Steroids
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications