Title of article :
Optimization of the trace element determination by ICP-MS in human blood serum
Author/Authors :
De Blas Bravo، نويسنده , , Isabel and Sanz Castro، نويسنده , , Ramَn and Lَpez Riquelme، نويسنده , , Natividad and Tormo Dيaz، نويسنده , , Consuelo and Apraiz Goyenaga، نويسنده , , David، نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2007
Pages :
4
From page :
14
To page :
17
Abstract :
The ICP-MS for simultaneous trace element determination in human blood has prevailed as the most suitable methodology for clinical aims because of its rapidity, detection limits and minimal sample quantity needed for analysis. proteic matrix is high, it is necessary to fine-tune the ICP-MS Agilent 7500i with autosampler CETAC ASX-500 and ISIS System connected, and further we have to the sample pre-treatment in order to obtain good results. udy of the results shows that the best pre-treatment for blood serum samples consists of a basic treatment by 1/10 dilution with a solution of EDTA and NH4OH, with a detection limit of the order of μg/L and a reduction of the necessary patient sample volume to 2 mL.
Keywords :
pre-treatment , optimization , Human blood serum , ICP-MS , trace element
Journal title :