Title of article
Separation and determination of Se-compounds by liquid chromatography coupled with electrospray mass spectrometry
Author/Authors
Carmen Barciela-Alonso، نويسنده , , M. and Bermejo-Barrera، نويسنده , , Adela and Bermejo-Barrera، نويسنده , , Pilar، نويسنده ,
Issue Information
فصلنامه با شماره پیاپی سال 2007
Pages
3
From page
23
To page
25
Abstract
A method for Selenocystine and Selenomethionine determination by LC–ES–MS was developed in this work. The mass spectrometer was used in a positive mode and the m/z used for the identification of Selenomethionine and Selenocystine were 198.35 and 337.15, respectively.
lenium species were separated using a LC system. A silica chromatographic column (ZORBAX Eclipse XDB-C8 of 50 mm length and 2.1 mm internal diameter (particle size 3.5 μm)) was used. The separation was realised in isocratic mode, using methanol:water (1:1) with 1% of acetic acid and a flow rate of 200 μL min−1. The developed method was precise (RSD of 4.5% and 3.9% for Selenomethionine and Selenocystine, respectively) and sensible (limit of detection (LOD) 0.06 and 0.99 mg L−1 for selenomethionine and selenocystine, respectively).
Keywords
Seleno-compounds , Liquid chromatography , mass spectrometry
Journal title
Journal of Trace Elements in Medicine and Biology
Serial Year
2007
Journal title
Journal of Trace Elements in Medicine and Biology
Record number
1724552
Link To Document