Title of article :
Measurement of ejection fraction with standard thermodilution catheters
Author/Authors :
dos Santos، نويسنده , , Icaro and Ferreira da Rocha، نويسنده , , Adson and Assis de O.Nascimento، نويسنده , , Francisco and Souza Neto، نويسنده , , Joao and Valvano، نويسنده , , Jonathan W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Right ventricle ejection fraction (RVEF) is clinically used to evaluate right ventricular function. The thermodilution method can be modified to estimate the RVEF. However, this method requires a thermistor with a fast time response in order to yield correct estimates. Digital signal processing techniques that were developed in previous works, allow the use of industry-standard slow time response thermistors for the measurement EF. However, these algorithms were not automated, and the works did not present a complete evaluation of the method’s performance. This article presents a modified automated version of these algorithms, and uses numerical and in vitro simulations to test their performance. In the simulations, the measured ejection fraction was compared to the true ejection fraction. RVEFs ranging from 0.20 to 0.80 were tested for heart rates ranging from 30 to 120 heart beats per min. Statistical analysis of data showed that the new method presents an improved performance.
Keywords :
neural network , digital signal processing , EJECTION FRACTION , Thermodilution , thermistor
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics