Title of article :
An improved approach for measurement of stroke volume during treadmill exercise
Author/Authors :
Song، نويسنده , , C.G. and Kim، نويسنده , , D.W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
517
To page :
522
Abstract :
In this study, a new ensemble averaging technique was developed to measure cardiac output during treadmill exercise. Each dz/dt peak is used as a starting point for ensemble averaging, instead of the conventionally used R point of ECG, in order to prevent the peak dz/dt waveform (C point) to be smoothed, which results in increased dz/dtmax. When using the R point as a reference, time interval from the R point to the peak dz/dt varies for each heart beat. Signal to noise ratios (SNR) of the five subjects with the new technique were 12.9, 16. 4, 10.6, 14.0, and 4.6% higher than those with the conventional one at rest and during four levels of exercise, respectively. Power density spectra of the averaged dz/dt waveforms by the peak dz/dt and R point referencing were also analyzed. Stroke volumes of the five subjects averaged by the peak dz/dt were 4.9, 6.8, 7.6, 7.4, and 9.8% higher than those by the R point at rest and during four levels of exercise, respectively.
Keywords :
Stroke volume , Impedance cardiography , ensemble average
Journal title :
Medical Engineering and Physics
Serial Year :
2003
Journal title :
Medical Engineering and Physics
Record number :
1728029