Three-dimensional analysis of electrode behavior in a human cochlear model
Author/Authors
Lim، نويسنده , , Yoon Seob and Park، نويسنده , , Se-Ik and Kim، نويسنده , , Yong Hyup and Oh، نويسنده , , Seung Ha and Kim، نويسنده , , Sung June، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
9
From page
695
To page
703
Abstract
Three-dimensional (3D) finite element analysis is used in this study to model the mechanical effects of the electrode in a cochlear implant. We employ six electrodes with different stiffness produced by different arrangements of metal wires. Different wire arrangements are generated by changing the fitness function of a genetic program. The human cochlea is modeled by the spiral-approximation method. Reconstructed three-dimensional CT images are used to model the real insertion condition. The contact pressure at the tip and the insertion force are found to be highest when the wires are stacked horizontally. Axial rotation of the electrode has minimal effect on the stimulating current spread. The electrode does not contact the basilar membrane. The results indicate that considering the electrode stiffness is important to minimizing mechanical trauma in cochlear implantation.
Keywords
Finite element analysis , Cochlear electrode , cochlear implant