Title of article
Validation of a digital image processing software package for the in vivo measurement of wear in cemented Charnley total hip arthroplasties
Author/Authors
Kennard، نويسنده , , Emma and Wilcox، نويسنده , , Ruth K. and Hall، نويسنده , , Richard M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
7
From page
356
To page
362
Abstract
Computer-generated images were used to assess image processing software employed in the radiographic evaluation of penetration in total hip replacement. The images were corrupted using Laplacian noise and smoothed to simulate different modulation transfer functions in a range associated with hospital digital radiographic systems. With no corruption, the penetration depth measurements were both precise and accurate. However, as the noise increased so did the inaccuracy and imprecision to levels that may make changes in the penetration observed clinically difficult to discern between follow-up assessments. Simulated rotation of the wire marker produced significant bias in the measured penetration depth. The use of these simulated radiographs allows the evaluation of the software used to process the digital images alone rather than the whole measurement system.
Keywords
Imaging , Total hip replacement , WEAR , Radiography
Journal title
Medical Engineering and Physics
Serial Year
2006
Journal title
Medical Engineering and Physics
Record number
1729036
Link To Document