Title of article :
Modeling, analysis, and validation of a pneumatically driven left ventricle for use in mock circulatory systems
Author/Authors :
Colacino، نويسنده , , F.M. and Arabia، نويسنده , , M. and Moscato، نويسنده , , F. and Danieli، نويسنده , , G.A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
11
From page :
829
To page :
839
Abstract :
In this work a pneumatically driven mock left ventricle and a native left ventricle are modeled and alternatively connected to the numerical model of a closed circulatory system comprising the systemic circulation, the left atrium, and inlet and outlet ventricular valves. By simulating physiological changes of the system working conditions, behavior and preload sensitivity of the pneumatic ventricle have been compared to those of a native ventricle. s show that a pneumatic ventricle, when used as a fluid actuator in mock circulations, has low flexibility in reproducing different scenarios, its interaction with peripheral circulatory districts is characterized by non-physiological values, and its preload sensitivity is in poor agreement with physiological data. Results’ analysis also shows that present mock circulatory systems for testing cardiovascular prostheses are inadequate, if a careful attention is not paid to the pumping action of the pneumatic ventricle. esented computer model, validated by comparing numerical results with in vitro measurements available in the literature, can be used for designing in vitro experiments, while choosing the best control strategy for pneumatic systems.
Keywords :
Mock circulatory system , Non-linear elastance , Test Bench , Pneumatic drive line , starling , cardiac output , MODELING , Venous return , Pneumatic ventricle
Journal title :
Medical Engineering and Physics
Serial Year :
2007
Journal title :
Medical Engineering and Physics
Record number :
1729550