Title of article :
Simulated parametric studies in optical imaging of tumors through temporal log-slope difference mapping
Author/Authors :
Guo، نويسنده , , Zhixiong and Wan، نويسنده , , Siew Kan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Numerical studies are conducted to evaluate the noninvasive imaging method of temporal log-slope difference mapping (TLSDM). Emphasis is placed on the parametric examination of tumorʹs size, enhanced absorption contrast ratio, and embedment depth on the imaging quality and accuracy. It is found that the imaging quality and accuracy are predominantly determined by the tumor size and its embedment depth. The TLSDM method can detect a small tumor of 10 mm in size when the tumor is embedded at a shallow depth down to 10 mm. With decreasing tumor size and/or increasing embedment depth, both image quality and accuracy worsen. The imaging method requires the enhanced absorption contrast ratio be 20:1 or above.
Keywords :
Tumor detection , Absorption contrast , Optical imaging , Temporal log-slope
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics