Title of article :
A low-cost thermoelectrically cooled tissue clamp for in vitro cyclic loading and load-to-failure testing of muscles and tendons
Author/Authors :
Kiss، نويسنده , , Marc-Olivier and Hagemeister، نويسنده , , Nicola and Levasseur، نويسنده , , Annie and Fernandes، نويسنده , , Julio and Lussier، نويسنده , , Bertrand and Petit، نويسنده , , Yvan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
In vitro cyclic loading and load-to-failure testing of muscles and tendons require a reliable linking device between tissues and the actuator that can transmit high loads without slippage or tissue damage. This article describes a simple custom-made thermoelectrically cooled freeze clamp. The effectiveness of the clamp to transmit loads without tissue slippage was evaluated on 10 canine quadriceps femoris myo-tendinous junctions in both load-to-failure and cyclic loading settings. Dynamic cyclic loading during an extensive period of time was successfully achieved. Loads up to 4.84 kN were applied in quasi-static conditions without evidence of clamp slippage or failure.
Keywords :
Freeze clamp , Tissue clamp , Static Loading , Cyclic loading
Journal title :
Medical Engineering and Physics
Journal title :