• Title of article

    Vibrotactile perception thresholds at the sole of the foot: Effects of contact force and probe indentation

  • Author/Authors

    Gu، نويسنده , , Cheng and Griffin، نويسنده , , Michael J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    6
  • From page
    447
  • To page
    452
  • Abstract
    When using vibrotactile thresholds to investigate neuropathy in the fingers, the indentation of a vibrating probe, and the force applied to a static surround around a vibrating probe, affect thresholds. This study was designed to investigate the effects on vibrotactile perception thresholds at the sole of the foot of probe indentation (i.e. height of a vibrating probe relative to a static surround) and the force applied to the static surround. Thresholds at 20 Hz (expected to be mediated by the NP I channel) and at 160 Hz (expected to be mediated by the Pacinian channel) were obtained at the hallux (i.e. greater toe) and the ball of the foot on 14 healthy subjects. In one condition, the height of the vibrating probe was varied to 0, 1, 2, 3, and 4 mm above a static surround with 4-N force applied to the surround. In a second condition, the force applied to the surround was varied to 1, 2, 3, 4, 5, and 6 N while using a probe height of 1 mm. Thresholds at 20 Hz decreased with increasing probe height from 0 to 1 mm but showed no significant variation between 2, 3, and 4 mm at either the hallux or the ball of the foot. Thresholds at 160 Hz decreased with increasing probe height from 0 to 4 mm at both the hallux and the ball of the foot. Thresholds at 20 Hz obtained with 1-N surround force were higher than thresholds obtained with 2 N, but there was no significant difference with surround forces from 2 to 6 N at either the hallux or the ball of the foot. Thresholds at 160 Hz were unaffected by variations in surround force at the ball of the foot but tended to decrease with increasing force at the hallux. It is concluded that a vibrating probe flush with a static surround, and a surround force in the range 2–4 N, are appropriate when measuring vibrotactile thresholds at the hallux and the ball of the foot with a 6-mm diameter contactor and a 2-mm gap to the static surround.
  • Keywords
    Toe , Foot , Frequency , Neuropathy , Vibration , Thresholds , Force , Indentation
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2012
  • Journal title
    Medical Engineering and Physics
  • Record number

    1731610