Title of article :
A non-invasive Cochlear Microphonic measurement system
Author/Authors :
Masood، نويسنده , , Asim and Teal، نويسنده , , Paul D. and Hollitt، نويسنده , , Christopher، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
The Cochlear Microphonic is one of the electrical potentials generated by the ear in response to audible stimuli. It is very difficult to measure the CM non-invasively because it has a very small magnitude (less than 1 μV). A high Common Mode Rejection Ratio (CMRR) and very large bandwidth (5 Hz–20 kHz) biomedical amplifier system is presented to measure the signal. The system also uses a driven right leg circuit to increase the CMRR.
Keywords :
averaging , filtering , shielding , Common Mode Rejection Ratio , Driven right leg circuit , cochlear microphonic
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics