Title of article
Needle-free jet injection using real-time controlled linear Lorentz-force actuators
Author/Authors
Taberner، نويسنده , , Andrew and Hogan، نويسنده , , N. Catherine and Hunter، نويسنده , , Ian W.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
8
From page
1228
To page
1235
Abstract
Needle-free drug delivery by jet injection is achieved by ejecting a liquid drug through a narrow orifice at high pressure, thereby creating a fine high-speed fluid jet that can readily penetrate skin and tissue. Until very recently, all jet injectors utilized force- and pressure-generating principles that progress injection in an uncontrolled manner with limited ability to regulate delivery volume and injection depth. In order to address these shortcomings, we have developed a controllable jet injection device, based on a custom high-stroke linear Lorentz-force motor that is feed-back controlled during the time-course of an injection.
this device, we are able to monitor and modulate continuously the speed of the drug jet, and regulate precisely the volume of drug delivered during the injection process. We demonstrate our ability to control injection depth (up to 16 mm) and repeatably and precisely inject volumes of up to 250 μL into transparent gels and post-mortem animal tissue.
Keywords
Lorentz-force motor , Jet-injection , Drug-delivery , Needle-free , Real-time Control