Title of article :
Individual-specific tonometry on porcine eyes
Author/Authors :
Leung، نويسنده , , Leo K.K. and Ko، نويسنده , , Match W.L. and Lam، نويسنده , , David C.C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
Intraocular pressure (IOP) monitoring is important in the diagnosis and management of glaucoma. The measurement of IOP is affected by corneal properties, but the effect of corneal stiffness on IOP measurement is unaccounted for in pressure measurement instruments such as the Goldmann Applanation Tonometer (GAT). A new instrumented non-invasive indentation tonometry that can measure IOPIST, a corneal stiffness-corrected intraocular pressure is developed. The inter-individual corneal variations of 12 porcine eyes ex vivo were independently characterized; and their true intraocular pressure, IOPTʹs, were set using a manometer before indentation using the new indentation tonometry. Analyses of the load–displacement data showed that porcine corneal stiffness varied more than five times from 0.045 to 0.253 N/mm. Analysis showed that, without individual stiffness correction, inter-individual variation of IOPGAT can vary up to 8 mm Hg from IOPT at 15 mm Hg; the error becomes larger at high IOPT. In comparison when corneal stiffness is accounted for, IOPIST has a significantly smaller error of 1.82 ± 1.70 mm Hg for IOPT between 12 and 40 mm Hg than IOPGAT. The results showed that the new tonometry successfully accounted for inter-individual variations in IOP measurement.
Keywords :
Intraocular pressure , Biomechanics , cornea , Stiffness , Glaucoma , Tonometry
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics