Title of article :
Development and Validation of a Novel Method to Derive Central Aortic Systolic Pressure From the Radial Pressure Waveform Using an N-Point Moving Average Method
Author/Authors :
Williams، نويسنده , , Bryan and Lacy، نويسنده , , Peter S. and Yan، نويسنده , , Peter and Hwee، نويسنده , , Chua-Ngak and Liang، نويسنده , , Chen and Ting، نويسنده , , Lee Choon Meng، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
11
From page :
951
To page :
961
Abstract :
Objectives rpose of this study was to develop and validate the novel application of a simple n-point moving average (NPMA)—a mathematical low pass filter—to noninvasively derive central aortic systolic pressure (CASP) from the radial artery pressure waveform (RAPWF) in humans. ound ay be an important independent determinant of clinical outcomes. Development of simple, well-validated methods to noninvasively derive CASP is necessary to facilitate the routine clinical measurement of CASP. s studies in different population cohorts were used to develop and validate the NPMA method to derive CASP in humans: 1) a development study (n = 217), describing the optimal application of the NPMA to derive CASP; 2) a validation study comparing NPMA-CASP with CASP derived using a generalized transfer function (GTF-CASP [SphygmoCor system, AtCor, Sydney, Australia]) using 5,349 RAPWFs from the CAFE (Conduit Artery Function Evaluation) study; and 3) an invasive validation study (n = 20) comparing NPMA-CASP with direct aortic root pressure measurements during cardiac catheterization. s development study, when using the NPMA, a denominator of n/4 (where n = tonometer sampling frequency) most accurately defined CASP relative to GTF-CASP. Validation of NPMA-CASP using RAPWFs from the CAFE study revealed excellent correlation and agreement (r2 = 0.993, mean difference 0.3 ± 1.0 mm Hg). The agreement remained robust after stratification by sex, age, treatment, and diabetes status. There was also excellent correlation and agreement (r2 = 0.98, p < 0.001) between directly measured aortic root systolic pressures (Millarʹs catheter) at cardiac catheterization versus NPMA-CASP, derived simultaneously from noninvasive RAPWFs. sions w that an NPMA with a denominator of one-quarter of the tonometer sampling frequency accurately defines CASP when applied to noninvasively acquired RAPWFs in man. These novel findings have important implications for the simplification of noninvasive CASP measurement and its wider application in clinical trials and clinical practice.
Keywords :
central aortic systolic pressure , Noninvasive , n-point moving average , radial artery pressure waveforms , Generalized transfer function
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
2011
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
1751524
Link To Document :
بازگشت