Title of article :
Percutaneous Left Atrial Appendage Closure With the AMPLATZER Cardiac Plug Device in Patients With Nonvalvular Atrial Fibrillation and Contraindications to Anticoagulation Therapy
Author/Authors :
Urena، نويسنده , , Marina and Rodés-Cabau، نويسنده , , Josep and Freixa، نويسنده , , Xavier and Saw، نويسنده , , Jacqueline and Webb، نويسنده , , John G. and Freeman، نويسنده , , Mélanie and Horlick، نويسنده , , Eric and Osten، نويسنده , , Mark and Chan، نويسنده , , Albert and Marquis، نويسنده , , Jean-Francois and Champagne، نويسنده , , Jean and Ibrahim، نويسنده , , Réda، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
Objectives
m of this study was to evaluate the results associated with left atrial appendage closure (LAAC) with the AMPLATZER Cardiac Plug (ACP) (St. Jude Medical, Minneapolis, Minnesota) in patients with nonvalvular atrial fibrillation and absolute contraindications to anticoagulation therapy.
ound
ta exist on the late outcomes after LAAC in patients with absolute contraindications to warfarin.
s
l of 52 patients with nonvalvular atrial fibrillation underwent LAAC with the ACP device in 7 Canadian centers. Most patients received short-term (1 to 3 months) dual-antiplatelet therapy after the procedure and single-antiplatelet therapy thereafter. A transesophageal echocardiography was performed in 74% of patients at the 6-month follow-up. No patient was lost to follow-up (≥12 months in all patients).
s
an age and median (interquartile range) CHADS2 score were 74 ± 8 years and 3 (2 to 4), respectively. The procedure was successful in 98.1% of the patients, and the main complications were device embolization (1.9%) and pericardial effusion (1.9%), with no cases of periprocedural stroke. At a mean follow-up of 20 ± 5 months, the rates of death, stroke, systemic embolism, pericardial effusion, and major bleeding were 5.8%, 1.9%, 0%, 1.9%, and 1.9%, respectively. The presence of mild peridevice leak was observed in 16.2% of patients at the 6-month follow-up as evaluated by transesophageal echocardiography. There were no cases of device thrombosis.
sions
ients with nonvalvular atrial fibrillation at high risk of cardioembolic events and absolute contraindications to anticoagulation, LAAC using the ACP device followed by dual-/single-antiplatelet therapy was associated with a low rate of embolic and bleeding events after a mean follow-up of 20 months. No cases of severe residual leak or device thrombosis were observed at the 6-month follow-up.
Keywords :
atrial fibrillation , percutaneous closure of left atrial appendage , AMPLATZER Cardiac Plug , Anticoagulation
Journal title :
JACC (Journal of the American College of Cardiology)
Journal title :
JACC (Journal of the American College of Cardiology)