Title of article :
Transcatheter Implantation of Homologous “Off-the-Shelf” Tissue-Engineered Heart Valves With Self-Repair Capacity: Long-Term Functionality and Rapid In Vivo Remodeling in Sheep
Author/Authors :
Petra E. and Driessen-Mol، نويسنده , , Anita and Emmert، نويسنده , , Maximilian Y. and Dijkman، نويسنده , , Petra E. and Frese، نويسنده , , Laura E. Sanders، نويسنده , , Bart and Weber، نويسنده , , Benedikt and Cesarovic، نويسنده , , Nikola and Sidler، نويسنده , , Michele and Leenders، نويسنده , , Jori and Jenni، نويسنده , , Rolf and Grünenfelder، نويسنده , , Jürg and Falk، نويسنده , , Volkmar and Baaijens، نويسنده , , Frank P.T. and Hoerstrup، نويسنده , , Simon P.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
Objectives
tudy sought to evaluate long-term in vivo functionality, host cell repopulation, and remodeling of “off-the-shelf” tissue engineered transcatheter homologous heart valves.
ound
atheter valve implantation has emerged as a valid alternative to conventional surgery, in particular for elderly high-risk patients. However, currently used bioprosthetic transcatheter valves are prone to progressive dysfunctional degeneration, limiting their use in younger patients. To overcome these limitations, the concept of tissue engineered heart valves with self-repair capacity has been introduced as next-generation technology.
s
o functionality, host cell repopulation, and matrix remodeling of homologous transcatheter tissue-engineered heart valves (TEHVs) was evaluated up to 24 weeks as pulmonary valve replacements (transapical access) in sheep (n = 12). As a control, tissue composition and structure were analyzed in identical not implanted TEHVs (n = 5).
s
atheter implantation was successful in all animals. Valve functionality was excellent displaying sufficient leaflet motion and coaptation with only minor paravalvular leakage in some animals. Mild central regurgitation was detected after 8 weeks, increasing to moderate after 24 weeks, correlating to a compromised leaflet coaptation. Mean and peak transvalvular pressure gradients were 4.4 ± 1.6 mm Hg and 9.7 ± 3.0 mm Hg, respectively. Significant matrix remodeling was observed in the entire valve and corresponded with the rate of host cell repopulation.
sions
e first time, the feasibility and long-term functionality of transcatheter-based homologous off-the-shelf tissue engineered heart valves are demonstrated in a relevant pre-clinical model. Such engineered heart valves may represent an interesting alternative to current prostheses because of their rapid cellular repopulation, tissue remodeling, and therewith self-repair capacity. The concept of homologous off-the-shelf tissue engineered heart valves may therefore substantially simplify previous tissue engineering concepts toward clinical translation.
Keywords :
transcatheter valve implantation , off-the-shelf tissue engineering , tissue-engineered heart valves , Self-repair