AbstractBackground
er ablation is important for treatment of paroxysmal atrial fibrillation (PAF). Limited animal and human studies suggest a correlation between electrode-tissue contact and radiofrequency lesion generation.
ives
udy sought to assess the safety and effectiveness of an irrigated, contact force (CF)-sensing catheter in the treatment of drug refractory symptomatic PAF.
s
pective, multicenter, nonrandomized study was conducted. Enrollment criteria included: ≥3 symptomatic episodes of PAF within 6 months of enrollment and failure of ≥1 antiarrhythmic drug (Class I to IV). Ablation included pulmonary vein isolation with confirmed entrance block as procedural endpoint.
s
l of 172 patients were enrolled at 21 sites, where 161 patients had a study catheter inserted and 160 patients underwent radiofrequency application. Procedural-related serious adverse events occurring within 7 days of the procedure included tamponade (n = 4), pericarditis (n = 3), heart block (n = 1, prior to radiofrequency application), and vascular access complications (n = 4). By Kaplan-Meier analyses, 12-month freedom from atrial fibrillation/atrial flutter/atrial tachycardia recurrence was 72.5%. The average CF per procedure was 17.9 ± 9.4 g. When the CF employed was between investigator selected working ranges ≥80% of the time during therapy, outcomes were 4.25 times more likely to be successful (p = 0.0054; 95% confidence interval: 1.53 to 11.79).
sions
ART-AF trial demonstrated that this irrigated CF-sensing catheter is safe and effective for the treatment of drug refractory symptomatic PAF, with no unanticipated device-related adverse events. The increased percent of time within investigator-targeted CF ranges correlates with increased freedom from arrhythmia recurrence. Stable CF during radiofrequency application increases the likelihood of 12-month success. (THERMOCOOL® SMARTTOUCH® Catheter for Treatment of Symptomatic Paroxysmal Atrial Fibrillation; NCT01385202)