Title of article :
Multicenter Evaluation of a Next-Generation Balloon-Expandable Transcatheter Aortic Valve
Author/Authors :
Webb، نويسنده , , John and Gerosa، نويسنده , , Gino and Lefèvre، نويسنده , , Thierry and Leipsic، نويسنده , , Jonathon and Spence، نويسنده , , Mark and Thomas، نويسنده , , Martyn and Thielmann، نويسنده , , Matthias and Treede، نويسنده , , Hendrik and Wendler، نويسنده , , Olaf and Walther، نويسنده , , Thomas، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
AbstractBackground
PIEN 3 (Edwards Lifesciences Inc., Irvine, California) transcatheter valve incorporates features designed to address the well-known deficiencies of transcatheter aortic valve replacement (TAVR). An ultra–low-profile delivery system facilitates safe, controlled, and accurate implantation and an external seal minimizes paravalvular regurgitation.
ives
udy evaluated whether TAVR with this third-generation valve would be a viable alternative to high- or intermediate-risk surgery for severe aortic stenosis.
s
ospective study enrolled 150 patients at 16 sites in Europe and Canada. Clinical and echocardiographic outcomes were assessed at baseline, post-procedure, and 30 days. New sizing recommendations were developed during the course of the study.
s
ts were 83.6 ± 5.0 years of age, with multiple comorbidities reflected by a Society of Thoracic Surgeons score of 7.4 ± 4.5% and logistic EuroSCORE of 21.6 ± 12.3%. A transfemoral approach was chosen in 64.0% and alternative access (transapical/direct aortic) in the remainder. At 30 days, paravalvular regurgitation was none to mild in 96.4% and moderate in 3.5%. No patient had severe regurgitation. Transfemoral implantation was associated with low mortality (2.1%), no disabling stroke (0.0%), and fully percutaneous access and closure in 95.8%. Nontransfemoral alternative access was associated with higher rates of mortality (11.6%) and stroke (5.6%).
sions
hird-generation device addresses major deficiencies of earlier valves in terms of ease of use, accuracy of positioning, and paravalvular sealing. The rates of mortality and stroke with transfemoral access are among the lowest reported and support further evaluation as an alternative to open surgery in intermediate-risk patients. (Safety and Performance Study of the Edwards SAPIEN 3 Transcatheter Heart Valve [SAPIEN3]; NCT01808287)
Keywords :
Aortic stenosis , transcatheter aortic valve replacement
Journal title :
JACC (Journal of the American College of Cardiology)
Journal title :
JACC (Journal of the American College of Cardiology)