Author/Authors :
Verhaegen، Frank نويسنده , , Palmans، Hugo نويسنده ,
Abstract :
Current dosimetry protocols for clinical protons do not take into account any secondary electron fluence perturbation in ion chambers. In this work, we performed a systematic study of secondary electron fluence perturbation factors for spherical and cylindrical ion chambers in proton beams (70-250 MeV). The electron fluence perturbation factor, pe, was calculated using Monte Carlo transport of protons and secondary electrons. The influence of proton energy, cavity wall material (graphite, water, A150, PMMA, polystyrene), cavity radius, cavity wall thickness and positioning depth in water is studied. The influence of inelastic nuclear proton interactions is briefly discussed. It was found that pe depends on wall material; the largest values for pe were obtained for ion chambers with A150 walls (pe = 1.009), the smallest values for graphite walls. The perturbation factor was found to be largely independent of proton energy. A slight decrease of pe with cavity radius was obtained, especially for low energy protons. The wall thickness was found to have no effect on pe in the range studied (0.025-0.1 cm). The depth of the cavity in a water phantom was also found to have an insignificant effect on pe. Based on the results in the paper for spherical and cylindrical ion chambers, a method to calculate pe for a thimble ion chamber is presented. The results presented in this paper for cylindrical and spherical ion chambers are in contradiction to the calculated electron fluence perturbation factors for planar ion chambers in the paper by Casnati et al.