Abstract :
Strong transverse magnetic fields can produce very large dose enhancements and reductions in localized regions of a patient under irradiation by a photon beam. Through EGS4 Monte Carlo simulations, we have examined the effects of applying a magnetic field produced by a pair of circular current loops to a photon beam penetrating a water phantom of finite thickness. We have indeed found very substantial localized dose enhancements, albeit with no corresponding dose reduction just distal to the region of dose enhancement. (However, dose reduction does occur near the distal end of the phantom.) We have also observed two phenomena to be concerned with, for this configuration: significant broadening of the penumbra close to the current loop, and narrowness of the enhanced dose region in a plane parallel to the planes of the loops. We have also examined the use of a single current loop to produce the magnetic field, and have found great asymmetry in the dose distribution; this asymmetry appears to make it impossible to treat with a single circular magnet a tumor of large dimension extending below the application surface.