• Title of article

    Bilateral comparison of high vacuum primary standards between the Korea Research Institutes of Standards and Science (KRISS) and the National Measurement Institute of Japan (NMIJ)

  • Author/Authors

    Hong، نويسنده , , S.S. and Lim، نويسنده , , J.Y. and Hirata، نويسنده , , M. and Akimichi، نويسنده , , H.، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2011
  • Pages
    3
  • From page
    34
  • To page
    36
  • Abstract
    The bilateral comparison of high vacuum primary standards between the Korea Research Institute of Standards and Science (KRISS) and the National Measurement Institute of Japan (NMIJ) was performed in the nominal pressure range 4.09 × 10−3 Pa to 1.22 × 10−2 Pa by using two spinning rotor gauges as transfer standards. The accommodation coefficients (σ) of the two rotors for N2 determined by the two laboratories agreed to within 0.17% for rotor 10 and 0.89% for rotor 11. This agreement is within the expanded uncertainties (k = 1) of the primary standards at the two laboratories.
  • Keywords
    Vacuum primary standard , Bilateral comparison , Spinning rotor gauge , Expanded uncertainty
  • Journal title
    Current Applied Physics
  • Serial Year
    2011
  • Journal title
    Current Applied Physics
  • Record number