Title of article :
Electron dosimetry based on the absorbed dose to water concept: A comparison of the AAPM TG-51 and DIN 6800-2 protocols
Author/Authors :
Dohm، Oliver S. نويسنده , , Christ، Gunter نويسنده , , Nüsslin، Fridtjof نويسنده , , Schüle، Edmund نويسنده , , Bruggmoser، Gregor نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
-2257
From page :
2258
To page :
0
Abstract :
The dosimetry protocols DIN 6800-2 and AAPM TG-51, both based on the absorbed dose to water concept, are compared in their theoretical background and in their application to electron dosimetry. The agreement and disagreement in correction factors and energy parameters used in both protocols will be shown and discussed. Measurements with three different types of ionization chambers were performed and evaluated according to both protocols. As a result the perturbation correction factor P (60)co wall for the Roos chamber was determined to 1.024 ± 0.5%.
Keywords :
Fault current limiter , transient over voltage , power quality , short circuit current
Journal title :
MEDICAL PHYSICS
Serial Year :
2001
Journal title :
MEDICAL PHYSICS
Record number :
1791
Link To Document :
بازگشت