Title of article :
Improvement of blood compatibility of artificial heart valves via titanium oxide film coated on low temperature isotropic carbon
Author/Authors :
Wang، نويسنده , , Xianghui and Zhang، نويسنده , , Feng and Li، نويسنده , , Changrong and Zheng، نويسنده , , Zhihong and Wang، نويسنده , , Xi and Liu، نويسنده , , Xianghuai and Chen، نويسنده , , Anqing and Jiang، نويسنده , , Zhengbin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
Titanium oxide film was coated on the surface of low temperature isotropic carbon (LTI-carbon) by ion beam enhanced deposition (IBED) to improve the blood compatibility of the LTI-carbon mechanical heart valve. The prepared film was found to be polycrystalline with Ti2+, Ti3+ and Ti4+ coexisting. This oxygen-lack n-type semiconductor film exhibits a better hemocompatibility than that of LTI-carbon in both in vitro and in vivo investigations. The reason is attributed to the surface energy and work function, which are two influencing factors on blood compatibility.
Keywords :
IN VIVO , Blood compatibility , Ion beam enhanced deposition , IN VITRO , Titanium oxide film
Journal title :
Surface and Coatings Technology
Journal title :
Surface and Coatings Technology