Title of article :
Image reconstruction of sequentially sampled echo-planar data
Author/Authors :
Hyder، نويسنده , , Fahmeed and Rothman، نويسنده , , Douglas L. and Blamire، نويسنده , , Andrew M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
For echo-planar imaging (EPI), failure to time-reverse alternate echoes results in aliasing in the image. We encountered image artifacts in EPI acquired on a system with sequential sampling. After examining the source of these image artifacts, we concluded that the artifacts were a result of the type of sampling method used in data acquisition and the way the time-reversal of alternate echoes was carried out prior to Fourier transformation. Two methods are demonstrated to obtain artifact-free EPI with sequential data sampling.
Keywords :
Image reconstruction , Simultaneous sampling , Sequential Sampling , Time-reversal , Gradient-echo , MAGNETIC RESONANCE IMAGING , Echo-planar imaging , k-Space trajectory
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging