Title of article :
An in vitro study at low field for MR guidance of a biopsy needle
Author/Authors :
Arbogast-Ravier، نويسنده , , S. and Gangi، نويسنده , , A. and Choquet، نويسنده , , P. and Brunot، نويسنده , , B. and Constantinesco، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
The aim of this study is to demonstrate in vitro that low field strength can be used for MR interventional procedures. At 0.1 T, we developed 3D, T1, and T2-weighted fast steady state sequences and measured the artifact created by an MR biopsy needle (22 gauge, 0.7 mm diameter) on a bicompartmental phantom. The artifact generated by the needle placed perpendicular to Bo did not exceed 3.6 mm for 3D steady state sequences and acquisition times reaching 1.5 s. Corresponding artifacts at higher field strength were 7.6 mm for 0.5 T and 8.6 mm for 1 T. Low field MRI proved to be potentially adaptable for the guidance of MR biopsy needles.
Keywords :
Low field , Fast MRI , biopsy
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging