Title of article :
Modified birdcage coils for targeted imaging
Author/Authors :
Gasson، نويسنده , , Julia and Summers، نويسنده , , Ian R. and Fry، نويسنده , , Martin E. and Vennart، نويسنده , , William، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
10
From page :
1003
To page :
1012
Abstract :
The design of a radiofrequency coil of U-shaped geometry, specifically intended to image individual knuckle joints of the hand, has been investigated. The coil geometry is that of a “split and opened” birdcage coll. The optimum leg positions for such a U-shaped coil have been theoretically predicted using a novel technique that considers both the signal-to-noise ratio and the homogeneity of the field-of-view. Two particular coils of this type have been constructed, each optimum for a different size of image space and assuming a different source of image noise (sampledominated or coil-dominated). The experimentally determined radiofrequency fields produced by the coils correspond well to theory. Either coil can be used to obtain good-quality, high-resolution (130 μm × 130 μm in-plane) images in vivo of the first, second, or fifth metacarpophalangeal joints of the hand, sites of particular interest in the study of arthritis.
Keywords :
Radiofrequency coil , Birdcage coil , MRI , Knuckle , Signal-to-noise , MCP