• Title of article

    Determination of saturation transfer parameters of human tissues in vivo

  • Author/Authors

    Kajander، نويسنده , , Sami and Komu، نويسنده , , Markku and Niemi، نويسنده , , Pekka and Kormano، نويسنده , , Martti، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    5
  • From page
    413
  • To page
    417
  • Abstract
    In order to study the applicability of magnetization transfer contrast (MTC) to tissue differentiation, the determination of the magnetization transfer (MT) parameters of normal tissues is necessary for the evaluation of pathological conditions. The time-dependent saturation transfer technique was used to investigate the observed magnetization transfer parameters in several human tissues in vivo at 0.1 T. The length of the off-resonance saturation pulse varied from 0 to 750 ms. The magnetization transfer contrast (MTC) was 0.71 in striated muscle, 0.49 in liver, 0.49 in renal cortex, and 0.50 in spleen. The observed magnetization transfer rates (Rwm) were 5.5 s−1 for muscle, 3.1 s−1 for liver, and 1.5 s−1 for both renal cortex and spleen. Our results indicate that measuring Rwm and possibly other relaxation parameters could be useful in tissue differentiation.
  • Keywords
    Magnetization transfer contrast , Low-strength imaging , MRI , Tissue characterization
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1996
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1827276