Title of article :
Trabecular bone characterization with low-field MRI
Author/Authors :
Florence and Guillot، نويسنده , , Geneviève، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
4
From page :
639
To page :
642
Abstract :
Multiple-echo (ME) T2 measurements were performed in vitro at 0.1 T on a series of human trabecular bone samples filled with water and compared to site-matched bone density measurements. Multiple-echo T2 is strongly correlated to bone density and allows distinction between samples of equivalent bone density. Multiple-echo T2 is also dependent on the trabecular orientation with respect to the direction of the main field.
Keywords :
trabecular bone , Low-field magnetic resonance imaging , Multiple-echo T2