• Title of article

    Calculation of electric fields induced near metal implants by magnetic resonance imaging switched-gradient magnetic fields

  • Author/Authors

    Buechler، نويسنده , , D.N. and Durney، نويسنده , , C.H. and Christensen، نويسنده , , D.A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    10
  • From page
    1157
  • To page
    1166
  • Abstract
    Electric (E) fields induced near metal implants by MRI switched-gradient magnetic fields are calculated by a new equivalent-circuit numerical technique. Induced E-field results are found for a metallic spinal-fusion implant consisting of two thin wires connected to the metallic case of a current generator as well as for its subsections: a bare U-shaped wire, an insulated U-shaped wire, a cut insulated wire, and a generator. The presence of the metallic implants perturbs the E field significantly. Near the ends of the bare U-shaped wire, the E field is 89.7 times larger than in the absence of the wire. The greatest E field concentration occurs near the ends of the cut insulated wire, where the E field is 196.7 times greater than in the absence of the wire. In all cases, the perturbation of the induced E field by the implanted wire is highly localized within a few diameters of the wire.
  • Keywords
    Magnetic Resonance Imaging , Metallic implants , Induced electric fields , Safety
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1997
  • Journal title
    Magnetic Resonance Imaging
  • Record number