Title of article
Dual surface coil with high-B1 homogeneity for deep organ MR imaging
Author/Authors
Tomanek، نويسنده , , Boguslaw and Ryner، نويسنده , , Lawrence and Hoult، نويسنده , , David I. and Kozlowski، نويسنده , , Piotr and Saunders، نويسنده , , John K.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
6
From page
1199
To page
1204
Abstract
The theory and construction of a dual surface coil which provides good B1 homogeneity and sensitivity in a defined volume of interest is described. The probe comprises two coaxial rings, of different diameters and in different planes, which carry opposing currents of different values. Current in the second ring compensates for the roll-off of the B1 field associated with a single surface coil. Coupling between the rings and a third matching ring is by mutual inductance only. A comparison to a traditional surface coil with practical application to pig brain imaging at a field strength of 7 Tesla is shown.
Keywords
MRI , Surface Coil
Journal title
Magnetic Resonance Imaging
Serial Year
1997
Journal title
Magnetic Resonance Imaging
Record number
1829568
Link To Document