• Title of article

    Dual surface coil with high-B1 homogeneity for deep organ MR imaging

  • Author/Authors

    Tomanek، نويسنده , , Boguslaw and Ryner، نويسنده , , Lawrence and Hoult، نويسنده , , David I. and Kozlowski، نويسنده , , Piotr and Saunders، نويسنده , , John K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    1199
  • To page
    1204
  • Abstract
    The theory and construction of a dual surface coil which provides good B1 homogeneity and sensitivity in a defined volume of interest is described. The probe comprises two coaxial rings, of different diameters and in different planes, which carry opposing currents of different values. Current in the second ring compensates for the roll-off of the B1 field associated with a single surface coil. Coupling between the rings and a third matching ring is by mutual inductance only. A comparison to a traditional surface coil with practical application to pig brain imaging at a field strength of 7 Tesla is shown.
  • Keywords
    MRI , Surface Coil
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1997
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1829568