Title of article :
A Simple Phantom to Locate the Origin of MRI Ghost Artefacts
Author/Authors :
Taylor، نويسنده , , N.J. and Doyle، نويسنده , , V.L. and Fox، نويسنده , , R.A. and Leach، نويسنده , , M.O.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
4
From page :
73
To page :
76
Abstract :
A simple pyramidal tube phantom has been designed to allow accurate location of a selected slice along the desired axis of a magnetic resonance scanner and to provide a check of the quality of slice selection. It is particularly helpful in locating the origin of any out-of-slice ghost artefacts. The geometry of the phantom allows the slice positions to be calculated readily.
Keywords :
Artefacts , MRI , quality assurance , Phantom
Journal title :
Magnetic Resonance Imaging
Serial Year :
1998
Journal title :
Magnetic Resonance Imaging
Record number :
1829601
Link To Document :
بازگشت